LC Parts & Consumables Analytical Light Source GC Parts & Consumables Synthesis and Purification General Instruments and Equipment Biotech-relatedCapillary tube and connector Ghost Peak Sniper Check Valve and Valve Core Reagent Waste Liquid Sealing Cap Guard Column and Filter Two-Way Connector Chromatography Mobile Phase Filter General Laboratory Consumables DWK Reagent BottleDeuterium Lamp for Instrument Manufacturing Replacement Deuterium Lamp for Instruments Deuterium Lamp Power Supply Hollow Cathode Lamp Photoionization Lamp (PID Lamp) 5W Xenon Flash Lamp Module FiberLight® Fiber Optic LampFused Silica Capillary Chromatography Column Ultra-Inert Capillary Chromatography Column High-Temperature Capillary Chromatography Column Gas Chromatography ConsumablesAIS Purification & Organic MS System Automated Peptide Synthesizer Thin Film Evaporator Synthesis Workstation Stirrer
07
2025
-
05
Detection of Eight Organochlorine Pesticide Residues in Ginseng Using Gas Chromatography
Organochlorine pesticides With its lipophilic properties and high stability, it tends to accumulate persistently in human adipose tissue. Its metabolic rate is slow, and its cumulative effects can trigger neurological disorders, immune dysfunction, and pose a clear carcinogenic risk. These pesticides, by blocking nerve signal transmission or disrupting hormone secretion, may lead to developmental abnormalities of the brain—especially in children—slowed reaction times, and endocrine disruptions. Organochlorines can remain in soil and water bodies for decades, gradually accumulating through the food chain and eventually reaching hazardous concentrations in higher organisms. Once accumulated in the human body, organochlorines can be passed on to infants via breast milk, while poultry can transfer them into eggs, thus posing intergenerational health threats.
Shanghai HanKing Selection HK-5 GC Column 8 organochlorine pesticide residues in ginseng were detected and compared with... Agilent HP-5 for test comparison 。
Chromatographic conditions
Column: HK-5 (30 m × 0.32 mm × 0.25 μm)
Detector: ECD Detector
Injection port temperature: 230℃ , non-split injection
Detector temperature: 300℃
The column temperature is programmed to increase gradually: Initial temperature: 60℃, hold for 0.3 minutes; then increase at a rate of 60℃ per minute to 170℃; next, increase at a rate of 10℃ per minute to 220℃, hold for 10 minutes; then increase at a rate of 1℃ per minute to 240℃, followed by an increase at a rate of 15℃ per minute to 280℃, and hold for 5 minutes.
Detection Spectrum
![]() |
![]() |
Conclusion
Select HK-5 GC Column The residues of eight organochlorine pesticides in ginseng were detected and met the required detection standards. Compared with Agilent HP-5, The theoretical plate numbers and separation performance of the chromatographic peaks are essentially the same.
Keywords:
Related News
Contact Us
Address: Building 2, 8th Floor, No. 248 Guanghua Road, Minhang District, Shanghai
Scan to follow us
Scan to follow us
Call us.
Copyright © 2026 Shanghai HanKing Instrument & Equipment Co.,Ltd
Terms of Use
Privacy Policy
Address: Building 2, 8th Floor, No. 248 Guanghua Road, Minhang District, Shanghai市
Scan to follow us
Scan to follow us
Call us.
Copyright © 2026 Shanghai HanKing Instrument & Equipment Co.,Ltd
Terms of Use
Privacy Policy
Technical Support: Zhongqi Cross-border
Hu (Min) Emergency Management Crisis Xu [2023]201905