LC Parts & Consumables Analytical Light Source GC Parts & Consumables Synthesis and Purification General Instruments and Equipment Biotech-relatedCapillary tube and connector Ghost Peak Sniper Check Valve and Valve Core Reagent Waste Liquid Sealing Cap Guard Column and Filter Two-Way Connector Chromatography Mobile Phase Filter General Laboratory Consumables DWK Reagent BottleDeuterium Lamp for Instrument Manufacturing Replacement Deuterium Lamp for Instruments Deuterium Lamp Power Supply Hollow Cathode Lamp Photoionization Lamp (PID Lamp) 5W Xenon Flash Lamp Module FiberLight® Fiber Optic LampFused Silica Capillary Chromatography Column Ultra-Inert Capillary Chromatography Column High-Temperature Capillary Chromatography Column Gas Chromatography ConsumablesAIS Purification & Organic MS System Automated Peptide Synthesizer Thin Film Evaporator Synthesis Workstation Stirrer
05
2024
-
12
Instructions for Setting Up and Using the HK086924 Deuterium Lamp
The software settings are as follows:
Step 1 : Click the deuterium lamp module on the operation panel with the mouse, and the following interface will appear, as shown in Figure 1-1.

Step 2: Tap Control to enter the following interface, as shown in Figure 1-2. Please select. Manual mode – the lamp type is DAD lamp. 。

Instructions for Testing the Intensity of Agilent Deuterium Lamps
Step 1: On the desktop, locate “Agilent Lab Advisor,” double-click to open it, and then connect the instrument.
Step 2: Click “Service and Diagnostics” and select the corresponding detector.
Step 3: Click “Wavelength Calibration” to perform wavelength calibration.
Step 4: After completing the wavelength calibration, click on “Intensity Test” to perform energy detection.

Special reminder: (When calibrating the wavelength with the DAD detector, the flow cell must be removed. Before calibrating the wavelength with the VWD detector, the flow cell needs to be cleaned and then replaced with pure water before proceeding with the wavelength calibration.)
New light usage time setting
Under LC1260, select “EMF” to enter the following interface. Click on the UV lamp-on time and the number of ignition attempts, then click the refresh button.

Test Report Example:

Keywords:
Related News
Contact Us
Address: Building 2, 8th Floor, No. 248 Guanghua Road, Minhang District, Shanghai
Scan to follow us
Scan to follow us
Call us.
Copyright © 2026 Shanghai HanKing Instrument & Equipment Co.,Ltd
Terms of Use
Privacy Policy
Address: Building 2, 8th Floor, No. 248 Guanghua Road, Minhang District, Shanghai市
Scan to follow us
Scan to follow us
Call us.
Copyright © 2026 Shanghai HanKing Instrument & Equipment Co.,Ltd
Terms of Use
Privacy Policy
Technical Support: Zhongqi Cross-border
Hu (Min) Emergency Management Crisis Xu [2023]201905